Specifications include, but are not limited to: Thermomechanical Operational * Temperature range (max) of -150 to 1,000C ____________________ (Optional) * Temperature precision of 1C * Furnace air cooling of 500+ <10 min * Maximum solid sample size of 26 mm (L) x 10 mm (D) * Maximum sample size for film/fiber with static operation 26 mm (L) x 1.0 mm (T) x 4.7 mm (W) * Maximum sample size for film/fiber with dynamic operation 26 mm (L) x .35 mm (T) x 4.7 mm (W) * Measurement precision of0.1% * Sensitivity of 15 nm * Resolution (digital) <0.15 nm * Dynamic baseline drift of <1 m (-100 to 500 C) * Force range of0.001 to 2 N * Force resolution of0.001 N * Frequency range of0.01 to 2 Hz * Ability to handle gas flow to yield a test atmosphere with inert, oxidizing, or reactive gases * Electrical: 120V * Operational Modes (standard, stress/strain, creep, stress relaxation, dynamic and modulated) Valid Responses: [Please Select], Yes, meets or exceeds, No, alternate specification