Specifications include, but are not limited to: 1 GXR-40 40 kW High Frequency Generator 10 To 400 Ma Range 40 TO 125 kVp, in 1kVp increments mAs Range: .1-500 Timer Range: 0.001-10 seconds, 38 steps Anatomical Programming: Self-Diagnostics, Anode heat Unit monitor, Error Messaging, History reporting log USB port Linearity/Reproducibility: Coefficient of linearity < 0.01 Coefficient of variation < 0.005 KvP Nominal Input power 208-240 VAC Single Phase 2 Floor Mounted Tubestand Tube rotation of +/-135º and column rotation of +/-90º for lateral exposures Longitudinal Rail Length/Travel 10 Feet Tube transverse Travel 8.6 inches (+/-) Tube mounted in enclosed housing device All cables enclosed in corrugated hose Electric Locks engaged when power is off Laser light mounted on handle bar for cassette centering to table and wall stand Must be floor mounted with no backings in wall required 3 Deluxe Four Way Float Top Table High quality precision bearing movement Tabletop length: 86.6”, width 29.5” Transverse travel: 10”, longitudinal travel: 39” Flat top design with low absorption material and patient weight capacity of 600 lbs. Recessed infrared light to release table top locks Twist lock connectors located in rear of table Grid cabinet: 17”x17” structure, locking in position when engaged Grid: 103 lines / inch, (40 lines/cm), 10:1 ratio, 34”-44” focal distance Heavy Duty Cassette Tray Must be floor mounted with no backings in wall required